Medthin™ 04 TiON
The latest addition to the Medthin™ coating portfolio
Medthin™ 04 was developed to inhibit bacterial colonization and growth. The Medthin™ 04 coating combines antimicrobial properties with high wear resistance, and chemical stability. This makes Medthin™ 04 an ideal coating to mitigate the spread of microbes on medical instruments as well as for applications around hospital use.
The coating´s antimicrobial properties and biocompatibility were successfully tested according to ISO 22196 and ISO 10993 (parts -5, -10, -11), respectively. The coating can be deposited at temperatures ranging between 70 ºC and 300 ºC, which allows for a variety of materials to be coated.

Measurements performed on witness coupons |
|
Material |
TiON |
Coating thickness |
1 - 5 μm |
Scratch resistance, LC3 [N] |
> 65 N |
Roughness, Ra |
≤ 0.3 μm |
Hardness HV 0.02 |
1000 - 1800 |
Coating temperature |
< 300 °C |
Service Temp. |
< 400 °C |
Friction vs. steel, dry |
< 0.4 |
Color |
light gold |
L*a*b* |
71/1/19 68/4/25 |