Medthin™ 04 TiON

TiON coating for use in hospital environment

Medthin 04 TiON, has “sand beach” color or gray color. The coating evidenced considerable reduction of colony-forming units (CFU) in testing according to ISO 22196 which makes it an ideal coating for use in hospital environment. The coating is deposited at 300°C but it can be adapted to the temperature of the base material per request.

Technical data

 

Material Description

Titanium Oxi-Nitride

Applicable on the following materials¹

all Stainless Steels, Titanium and Aluminium Alloys

Technologies of deposition

PVD

Colour

Gold or Grey

Thickness [µm]

1 to 5

Structure

Cubic, nanocrystalline to amorphous

Oxidation Resistance

Medium

Cleaning and Sterilization Resistance

Medium, all methods allowed

Antireflection

Medium

Electrical Conductivity

Poor

Antibacterial properties

>50% reduction in CFU

Nanohardness [GPa]

20 to 30

Young Modulus [GPa]

280 to 400

Friction coefficient²

0.4

Adhesion Scratch Test, LC3 [N]

>30N

Adhesion Rockwell or Mercedes test³

Class 1,2, HF1-3

Roughness, Ra (as coated) [nm]

~100

Corrosion resistance – Neutral Salt Spray

Max 30min

Biocompatibility data for instruments

Passed ISO 10993-5, ISO 10993-10, ISO 10993-11

MSDS

Delivered per request

MAF at FDA in the USA

none

Most popular applications

Color coding of instruments, cleaner surface

¹for any other material please contact ionbond
²against 100Cr6 steel ball
³Method either by ISO 26443 or VDI 3198

The values given are approximations and may vary depending on the application, environmental factors and test protocols.